Q-SUN Xe-2 operation error messages
Number |
Information |
Conditions & Suggested Measures |
M1 |
The chamber door is open |
The chamber door is open. |
M2 |
The lamp chamber door is open |
The lamp chamber door is open. If the lamp chamber door is already closed but this error occurs, check the interlock components of the lamp chamber door. |
M3 |
Warning: Chamber door interlock problem |
If the chamber door switches 1 and 2 are inconsistent, one of the interlock devices is damaged or needs adjustment. |
M10 |
Test completed |
If "Display Information Only" is selected, this message will appear when the test ends. The user needs to manually turn off the device. |
M11 |
Test completed |
If "Alarm" is selected, this message will appear when the test ends. An alarm will sound, but the device will continue to run. |
M12 |
Test completed, shutdown |
If "Stop" is selected, this message will appear when the test ends, and the equipment will shut down. |
M13 |
Test completed, shutdown |
If "Stop + Alarm" is selected, this message will appear when the test ends, and the equipment will shut down. |
M14 |
Please replace the lamp tube |
The lamp tube has been used for 3,000 hours. |
M15 |
Please perform routine maintenance - refer to the instruction manual |
The equipment has been in operation for 2,000 hours. For preventive maintenance information, refer to Section 16 of the product manual. |
M16 |
Blackboard temperature too high XXX℃ |
The blackboard temperature is 3°C or higher than the set value and remains so for 55 minutes. 1.The blackboard temperature setpoint is too low relative to the chamber temperature setpoint. |
M17 |
The blackboard temperature is too low: XXX℃ |
The blackboard temperature is 3°C or lower than the set value and remains so for 55 minutes. 1.The blackboard temperature setpoint is too high relative to the chamber temperature setpoint. |
M18 |
The chamber air temperature is too high: XXX℃ |
Please translate the following text into English. The chamber air temperature is 3°C or higher than the set value and remains so for 55 minutes. 1.The setpoint is too low relative to the laboratory temperature. |
M19 |
The chamber air temperature is too low: XXX℃ |
Please translate the following text into English: The chamber air temperature is 3°C or lower than the set value and remains so for 55 minutes. 1.The setpoint is too high relative to the laboratory temperature. |
M24 |
When an alarm occurs, the laboratory temperature is XXX℃ |
When this message appears, it is accompanied by one or more error messages. This information helps with diagnosis, as the laboratory temperature may be the cause of chamber temperature or humidity failures. |
M25 |
The temperature of the main board is too high: XX℃ |
The temperature of the main board is higher than the maximum allowable temperature. Move the Q-SUN to an air-conditioned room. |
M26 |
Fault in the lamp rectifier fan |
1.The fan is damaged. |
M27 |
Lamp tube/rectifier fan is on: should be off |
1.The fan relay is damaged. |
M30 |
Replace the battery |
The voltage of the main board battery is too low. Please replace the battery. |
M31 |
Calibrate the light sensor |
The lamp tube has been used for 500 hours. Please recalibrate the sensor. |
M33 |
Incorrect irradiance calibrator: UC20/340 should be used |
A 420 nm or TUV model is being used, but a 340 nm sensor is installed on the Q-SUN. |
M34 |
Incorrect irradiance calibrator: UC20/420 should be used |
A 340 nm or TUV model is used, but a 420 nm sensor is installed on the Q-SUN. |
M35 |
Incorrect irradiance calibrator: UC20/TUV should be used |
A 340 or 420 nm model is used, but a TUV sensor is installed on the Q-SUN. |
M43 |
Spray is off: should be on |
The Q-SUN is running the spray step, but the spray is off. Possible causes are: 1.Water supply is turned off. |
M44 |
Spray is on: should be off |
The Q-SUN is not running the spray step, but the spray is on. Possible causes are: 1.Solenoid valve is damaged. |
M46 |
Atomizer water level is too low |
The water level in the humidifier tray is too low. Check the water supply system; see Section 5.4 of the product manual for details. |
M49 |
Power interruption |
This message indicates that the power was disconnected and then restored during equipment operation. |
M50 |
The humidity in the chamber is XX% too high |
The humidity in the chamber is 5% or more higher than the set value. |
M51 |
The humidity in the chamber is XX% too low |
The humidity in the chamber is 5% or more lower than the set value. |
M52 |
The blackboard temperature deviates by XXX℃ |
The blackboard temperature is 10°C higher than the set value and persists for 1 minute. 1.The blackboard temperature set value is too low relative to the chamber air temperature. |
M53 |
The chamber air temperature deviates by XXX℃ |
The chamber air temperature is 10°C higher than the set value and persists for 1 minute. 1.The chamber air temperature set value is too low relative to the laboratory temperature. |
M56 |
The fan controller is damaged. |
1.Input voltage fluctuation. |
M58 |
The turntable is rotating: should stop |
Turntable relay failure. |
M59 |
The turntable is stopped: should rotate |
1.The turntable rotation sensor needs adjustment or is faulty. |
M60 |
Low irradiance: Chan x |
The lamp irradiance is 5% lower than the set value. Please replace the lamp and recalibrate. |
M61 |
Lamp failure: Chan x |
Irradiance is 30% lower than the set value. 1.The lamp is burned out. |
M63 |
Excessive irradiance: Chan x |
The irradiance is 5% higher than the set value. 1.Incorrect connection between the rectifier and the main board circuit. |
M64 |
The lamp is on: should be off |
The Q-SUN is running the dark step, but the main board detects irradiance higher than 0.05 W/m². Please replace the lamp relay. |
M65 |
The alternating current (AC) voltage exceeds the rated range |
This message indicates that the rectifier may have turned off due to low voltage. Check the Xe-2 power supply. |
M67 |
Lamp relay failure |
During the "relay self-check" process, the irradiance is higher than 0.05 W/m². 1.Lamp relay failure. |
M68 |
Operation relay failure |
During the "relay self-check" process, the irradiance is higher than 0.05 W/m². 1.Operation relay failure. |
M70 |
Blackboard temperature sensor failure XXX℃ |
The blackboard temperature is below 3°C or above 180°C. 1.Ensure the connector is properly connected. |
M72 |
Air temperature sensor failure |
The air temperature is below 3°C. 1.Ensure the connector is properly connected. |
M74 |
Laboratory temperature sensor failure |
The laboratory temperature is below 3°C. 1.Ensure the connector is properly connected. |
M77 |
Temperature and humidity sensor failure |
The reading of the temperature and humidity sensor is below 3%. 1.Ensure the connector is properly connected. |
M79 |
Exceeded maximum calibration time |
The calibration time is limited to two hours. If the calibration is not completed, please restart the Xe-2 and complete the calibration within two hours. |
M80 |
Flash failure |
The program checksum test has failed. Please contact Q-Lab Maintenance and Equipment Technical Support. |
M81 |
Flash data corruption |
The parameter and setting data are corrupted. Please contact Q-Lab Maintenance and Equipment Technical Support. |
M82 |
RAM damage, RAM reloading |
RAM data error. Please contact Q-Lab Maintenance and Equipment Technical Support. |
M90 |
Please replace the temperature and humidity sensor |
Please replace the temperature and humidity sensor. |
M101 |
SD card missing |
The SD card on the main board is missing or not properly installed. Please contact Q-Lab Maintenance. |
M103 |
System automatic reboot |
Q-SUN has automatically rebooted to avoid potential issues during operation. Please contact Q-Lab Maintenance. |
M104 |
The software has been successfully installed |
The software is successfully installed. |
M900 |
Communication failure of the status screen |
Communication between the motherboard and the status screen has been interrupted. The device will enter single-screen mode. |
M901 |
Communication failure of the main menu screen |
Communication between the main board and the main menu screen has been interrupted. The device will enter single-screen mode. |
M902 |
Communication failure of the motherboard |
This message will appear on the screen where error reports M900 or M901 have occurred. |
The above information has detailed step-by-step instructions and guidance in the random product manual. If you need the product manual or other assistance, please contact us!