Q-SUN Xe-3 operation error message
Number |
Information |
Conditions & Suggested Measures |
M1 |
The chamber door is open |
The chamber door is open. |
M2 |
The lamp chamber door is open |
The lamp chamber door is open. If this error occurs when the lamp chamber door is already closed, check the interlock components of the lamp chamber door. |
M10 |
Test completed |
If "Display only information" is selected, this message will appear when the test ends. The user needs to manually turn off the device. |
M11 |
Test completed |
If "Alarm" is selected, this message will appear when the test ends. An alarm will sound, but the device will continue to run. |
M12 |
Test completed, machine stopped |
If "Stop" is selected, this message will appear when the test ends. The device will stop operating. |
M13 |
Test completed, machine stopped |
If "Stop + Alarm" is selected, this message will appear when the test ends. The device will stop operating. |
M14 |
Please replace the lamp tube |
The lamp tube has been used for 3000 hours. |
M15 |
Please perform routine maintenance - refer to the instruction manual |
The device has been running for 2000 hours. For preventive maintenance information, refer to Section 15 of the product manual. |
M16 |
The blackboard temperature is too high: XXX℃ |
The blackboard temperature is 3 degrees Celsius or higher than the set value and remains so for 55 minutes. 1.The blackboard temperature setpoint is too low relative to the chamber temperature setpoint. |
M17 |
The blackboard temperature is too low: XXX℃ |
The blackboard temperature is 3 degrees Celsius or lower than the set value and persists for 55 minutes. 1.The blackboard temperature setpoint is too high relative to the chamber temperature setpoint. |
M18 |
The chamber air temperature is too high: XXX℃ |
The chamber air temperature is 3 degrees Celsius or higher than the set value and remains so for 55 minutes. 1.The setpoint is too low relative to the laboratory temperature. 4.The circulation fan drive is damaged. |
M19 |
The chamber air temperature is too low: XXX℃ |
The chamber air temperature is 3 degrees Celsius or lower than the set value and persists for 55 minutes. 1.The setpoint is too high relative to the laboratory temperature. |
M20 |
The chamber temperature is too high: XX℃ |
The chamber temperature is too high. 1.The air heater relay is damaged. |
M21 |
The chamber temperature is too low: XX℃ |
The chamber temperature is below the minimum allowable temperature. 1.The air heater relay is damaged. |
M24 |
Laboratory temperature during alarm: XXX℃ |
Laboratory temperature may be the cause of chamber temperature failure. |
M25 |
The main board temperature is too high: XX℃ |
The temperature of the main board is higher than the maximum allowable temperature. It is necessary to move the Q - SUN to an air - conditioned room. |
M26 |
Lamp rectifier fan failure
|
1.The fan is damaged. |
M27 |
Lamp/rectifier fan is on: should be off |
1.The fan relay is damaged. |
M28 |
Humidifier temperature is too high - may be empty: XXX℃ |
The humidifier temperature is too high. |
M29 |
Humidifier temperature is too low: XXX ℃ |
The humidifier temperature is too low. Check whether the humidifier water heater is working properly. |
M30 |
Replace the battery |
The voltage of the mainboard battery is too low. Please replace the battery. |
M31 |
Calibrate the light sensor |
The lamp has been used for 500 hours. Please recalibrate the sensor. |
M33 |
Incorrect irradiance calibrator: UC20/340 should be used |
The 420 nm or TUV model is used, but a 340 nm sensor is installed on the Q-SUN. Use the UC20/340 calibrator. |
M34 |
Incorrect irradiance calibrator: UC20/420 should be used |
The 340 nm or TUV model is used, but a 420 nm sensor is installed on the Q-SUN. Use the UC20/420 calibrator. |
M35 |
Incorrect irradiance calibrator: UC20/TUV should be used |
The 340 or 420 nm model is used, but a TUV sensor is installed on the Q-SUN. Use the UC20/TUV calibrator. |
M38 |
Light sensor mismatch |
The light sensors in the chamber are inconsistent. Please ensure all light sensors are the same. |
M43 |
Spray is off: should be on |
The Q-SUN is running the spray step, but the spray is off. Possible causes include: 1.Water supply is turned off. |
M44 |
Spray is on: should be off |
The Q-SUN is not running the spray step, but the spray is on. Possible causes are: 1.The solenoid valve is damaged. |
M45 |
Humidifier water level is too low |
1.The water supply is turned off. |
M49 |
Power interruption |
This message indicates that the power supply was interrupted and then restored during the equipment's operation. |
M50 |
Chamber humidity is too high: XX% |
The chamber humidity is 5% or more higher than the set value. |
M51 |
Chamber humidity is too low: XX% |
The chamber humidity is 5% or more lower than the set value. |
M52 |
Blackboard temperature deviation: XXX ℃ |
·The blackboard temperature is 10°C higher than the set value and persists for 1 minute. 1.The blackboard temperature set value is too low relative to the chamber air temperature. |
M53 |
The chamber air temperature deviates by XXX ℃ |
The chamber air temperature is 10°C higher than the set value and persists for 1 minute. 1.The chamber air temperature set value is too low relative to the laboratory temperature. |
M56 |
The fan controller is damaged. |
1.Input voltage fluctuation. |
M60 |
Low irradiance: Chan x |
The irradiance of the lamp is 5% lower than the set value. Please replace the lamp and recalibrate. |
M61 |
Lamp failure: Chan x |
The irradiance is 30% lower than the set value. 1.The lamp is burned out. |
M63 |
High irradiance: Chan x |
The irradiance is 5% higher than the set value. 1.The wiring between the rectifier and the mainboard is incorrect. |
M64 |
Lamp is on: should be off |
The Q-SUN is running the dark phase, but the mainboard detects irradiance above 0.05 W/m². Please replace the lamp relay. |
M65 |
AC voltage exceeds the rated range |
For 400V equipment, the input voltage is below 320V or above 460V, or the neutral wire is not connected. |
M67 |
Lamp relay malfunction |
During the "relay self-check" process, the irradiance is higher than 0.05 W/m². Please replace the lamp relay. |
M68 |
Operation relay malfunction |
During the "relay self-check" process, the irradiance is higher than 0.05 W/m². Please replace the operation relay. |
M69 |
Relay check in progress |
Displayed during M67 & M68 diagnostics. |
M70 |
Blackboard temperature sensor failure (XXX℃) |
The blackboard temperature is below 3°C. |
M72 |
Air temperature sensor failure |
The air temperature is below 3°C. |
M74 |
Laboratory temperature sensor failure |
The laboratory temperature is below 3°C. |
M76 |
Humidifier temperature sensor failure |
The humidifier temperature sensor reading is less than 3°C. Please replace the humidifier temperature sensor. |
M77 |
Temperature and humidity sensor failure |
The temperature and humidity sensor reading is below 3%. Please replace the temperature and humidity sensor. |
M80 |
Flash memory failure |
The program checksum test has failed. Please contact Q-Lab Maintenance and Equipment Technical Support. For contact information, refer to Section 19 of the product manual. |
M81 |
Flash memory data corruption |
The parameter and setting data are corrupted. Please contact Q-Lab Maintenance and Equipment Technical Support. For contact information, refer to Section 19 of the product manual. |
M82 |
RAM damaged, RAM reloading |
RAM data error. Please contact Q-Lab Maintenance and Equipment Technical Support. For contact information, refer to Section 19 of the product manual. |
M90 |
Please replace the temperature and humidity sensor |
The temperature and humidity sensor calibration has expired. Please replace the temperature and humidity sensor. |
M101 |
SD card missing |
The SD card on the mainboard is missing or not properly installed. Please contact Q-Lab Maintenance. For contact information, refer to Section 19 of the product manual. |
M103 |
System auto-reboot |
Q-SUN has automatically rebooted to avoid potential issues during operation. Please contact Q-Lab Maintenance. |
M104 |
Software installed successfully |
Software installed successfully. |
M900 |
Status screen communication failure |
Communication between the mainboard and the status screen has been interrupted. The equipment will enter single-screen mode. |
M901 |
Main menu screen communication failure |
Communication between the mainboard and the main menu screen has been interrupted. The equipment will enter single-screen mode. |
M902 |
Mainboard communication failure |
This message will be displayed on the screen where M900 or M901 error messages have occurred. |
The above information has detailed step-by-step instructions and guidance in the product manual provided with the device. For the product manual or other assistance, please contact us!