Q-SUN Xe-1 Operation Error Messages
Number |
Information |
Notes |
M1 |
CHAMBER DOOR IS OPEN |
The test chamber door is not properly closed. Please close the chamber door. If the M1 message still does not disappear after the chamber door is properly closed, the door interlock switch is faulty. |
M2 |
LAMP DOOR IS OPEN |
The door of the lamp chamber is not properly closed. Please close the chamber door. If the M2 message still does not disappear after the chamber door is properly closed, the door interlock switch is faulty. |
M5 |
BLACK PANEL SENSOR NOT CALIBRATED |
The blackboard temperature sensor is not calibrated. |
M7 |
AIR TEMP SENSOR NOT CALIBRATED |
The chamber temperature sensor is not calibrated. |
M10 |
END OF TEST |
The test is completed. This is just an information message. If "MESSAGE ONLY" is selected in the program, this message will appear when the test ends, and the test will continue. The user must manually stop the machine. |
M11 |
END OF TEST |
The test is completed. This is just an information message。 If "ALARM" is selected in the program, this message will appear when the test ends, and the test will continue. The user must stop the machine manually. |
M12 |
END OF TEST SHUTDOWN |
The test is completed. Power off. This is just an information message. If "STOP" is selected in the program, this message will appear when the test ends, and there will be no alarm sound. Automatic shutdown. |
M13 |
END OF TEST SHUTDOWN |
The test is complete. Power off. This is just an information message If "STOP+ALARM" is selected in the program, this message will appear when the test ends, and there will be an alarm sound. Automatic shutdown. |
M14 |
TIME TO REPLACE LAMPS |
Lamp replacement is required. This message will appear after the lamp has been used for 1,500 hours. Replace the lamp. |
M15 |
TIME FOR ROUTINE SERVICE- |
(After this message last appeared) The equipment has been running for 2,000 hours. Please refer to the maintenance content in Chapter 13. |
M16 |
BLACK PANEL TEMP TOO HOT XXX℃ |
The temperature of the blackboard temperature sensor is too low. This message appears when the sensor temperature is more than 3℃ below the set value and lasts for more than 55 minutes. Possible causes include: 1. The temperature set value of the blackboard temperature sensor is too high relative to the chamber temperature. 2. There is a problem with the circulating air fan or its drive. 3. There is a problem with the fresh air fan or its drive. Consult the Customer Service Department of Q-LAB Corporation. |
M17 |
BLACK PANEL TEMP TOO COLD XXX℃ |
The temperature of the blackboard temperature sensor is too low. This message appears when the sensor temperature is more than 3°C below the set value and lasts for more than 55 minutes. Possible causes include: 1.The temperature setpoint of the blackboard temperature sensor is too high relative to the chamber temperature. 2.There is a problem with the circulating air fan or its drive. 3.There is a problem with the fresh air fan or its drive. Consult the Customer Service Department of Q-LAB Corporation. |
M18 |
CHAMBER AIR TEMP TOO HOT XXX℃ |
The temperature of the chamber temperature sensor is too high. This message appears when the sensor temperature is more than 3°C above the set value and lasts for more than 55 minutes. Possible causes include: 1.The set value of the chamber temperature is too low relative to the laboratory temperature. 2.There is a problem with the heater relay. 3. There is a problem with the fresh air fan or its drive. 4. There is a problem with the circulating air fan or its drive. Consult the Customer Service Department of Q-LAB Corporation. |
M19 |
CHAMBER AIR TEMP TOO COLD XXX℃ |
The temperature of the chamber temperature sensor is too low. This message appears when the sensor temperature is more than 3°C below the set value and lasts for more than 55 minutes. Possible causes include: 1. The set value of the chamber temperature is too high relative to the laboratory temperature. 2. There is a problem with the heater relay. 3. There is a problem with the heater. 4. There is a problem with the drive of the fresh air fan. 5.There is a problem with the circulating air fan or its drive. Consult the Customer Service Department of Q-LAB Corporation. |
M20 |
CHAMBER TOO HOT XX℃ |
Possible causes of excessively high temperature in the test chamber include: 1. There is a problem with the heater relay. 2. There is a problem with the fan inside the chamber. Consult the Customer Service Department of Q-LAB Corporation. |
M21 |
CHAMBER TOO COLD XX℃ |
The temperature of the test chamber is lower than the minimum allowable temperature for the test. Possible causes include: 1. There is a problem with the heater relay. 2. There is a problem with the heater. 3. There is a problem with the fan inside the chamber. Consult the Customer Service Department of Q-LAB Corporation. |
M24 |
LAB TEMPERTURE AT ALARM XX°C |
The laboratory temperature may cause chamber temperature failures. |
M25 |
CONTROLLER TOO HOT XX℃ |
The temperature of the controller is too high. The temperature of the controller exceeds the maximum allowable temperature. Move the Q - Sun to an air - conditioned room. Consult the Customer Service Department of Q-LAB Corporation. |
M26 |
LAMP COOLING BLOWER FAILURE |
Tube cooling fan failure. Possible causes include: 1. There is a problem with the fan. 2. There is a problem with the fan relay. 3. There is a problem with the air flow switch. |
M27 |
LAMP COOLING BLOWER ON: SHOULD BE OFF |
The tube cooling fan is still running: it should stop. Possible causes include: 1.There is a problem with the fan relay. 2. There is a problem with the air flow switch. |
M28 |
HUMIDIFIER TOO HOT XXXC-MAY BE EMPTY |
The humidifier is too hot. It may be out of water. Check the water level. |
M29 |
HUMIDIFIER TOO COLD XXX℃ |
The temperature of the humidifier is too low. Check whether its heater is normal. |
M30 |
REPLACE BATTERY |
Replace the battery. The battery voltage in the control area is too low. Replace the battery. Note: Please dispose of lithium batteries in compliance with local regulations. |
M31 |
CALIBRATE LIGHT SENSORS |
The lamp has been used for 500 hours. Calibrate the sensor. |
M33 |
WRONG RADIOMETER:SHOULD BE CR20/340 |
The calibration instrument model is incorrect. A 420nm or TUV type was used, but a CR20/340 type calibration instrument should be used. |
M34 |
WRONG RADIOMETER:SHOULD BE CR20/420 |
The calibration instrument model is incorrect. A 340nm or TUV type was used, but a CR20/420 type calibration instrument should be used. |
M35 |
WRONG RADIOMETER:SHOULD BE CR20/TUV |
The calibration instrument model is incorrect. A 340 or 420 nm type was used, but a CR20/TUV type calibration instrument should be used. |
M38 |
LIGHT SENSORS DO NOT MATCH |
The light intensity sensors are incompatible. The light intensity sensors in the test chamber are not completely identical. It is necessary to ensure that all light intensity sensors are completely identical. |
M43 |
WATER SPRAY OFF:SHOULD BE ON |
Water spray stopped: it should be turned on. Possible causes include: 1. No water supply. 2. There is a problem with the solenoid valve. 3. There is a problem with the flow switch. |
M44 |
WATER SPRAY ON :SHOULD BE OFF |
Water spray is on: it should be stopped. Possible causes include: 1.There is a problem with the solenoid valve. 2. There is a problem with the flow switch. |
M45 |
HUMIDIFIER WATER LEVEL TOO LOW |
The water level in the humidifier is too low. Possible causes include: 1.No water supply. 2. There is a problem with the float switch. |
M50 |
CHAMBER HUMIDITY TOO HIGH XX% |
The humidity inside the chamber is too high. This message appears when the humidity is 5% higher than the set value. |
M51 |
CHAMBER HUMIDITY TOO LOW XX% |
The humidity inside the chamber is too low. This message appears when the humidity is 5% lower than the set value. |
M52 |
BLACK PANEL TEMP RUNAWAY XXX℃ |
The blackboard temperature is too high. The blackboard temperature exceeds the set value by 10 °C for more than 1 minute. Possible causes include: 1. The set value of the blackboard temperature is too low relative to the set value of the chamber temperature. 2. There is a problem with the circulating air fan or its drive. 3. There is a problem with the fresh air fan or its drive. Consult the Customer Service Department of Q-LAB Corporation. |
M53 |
CHAMBER AIR TEMP RUNAWAY XXX℃ |
The air temperature in the chamber is too high. The chamber air temperature exceeds the set value by 10°C for more than 1 minute. Possible causes include: 1. The set value is too low relative to the laboratory temperature. 2. There is a problem with the heater relay. 3. There is a problem with the fresh air fan or its drive. 4. There is a problem with the circulating air fan or its drive. Consult the Customer Service Department of Q-LAB Corporation. |
M54 |
CHAMBER TEMP RUNAWAY XXX℃ |
The chamber temperature is too high. The chamber temperature exceeds the set value by 10°C for more than 1 minute. Possible causes include: 1.The set value is too low relative to the laboratory temperature. Consult the Customer Service Department of Q-LAB Corporation. |
M56 |
BLOWER CONTROLLER FAULT |
The fresh air fan is not working because its controller is faulty. Possible causes include: 1.Unstable input voltage. Consult the Customer Service Department of Q-LAB Corporation. |
M60 |
LOW IRRADIANCECHANGE LAMP |
The light intensity is too low. The light intensity is 5% lower than the set value. Replace the lamp and recalibrate. |
M61 |
LAMP OUT |
Lamp failure. The light intensity is 30% lower than the set value. Possible causes include: 1.The lamp is burned out. |
M63 |
IRRADIANCE TOO HIGH |
The light intensity is too high. The light intensity is 5% higher than the set value. Possible causes include: 1.The dimming signal cable between the ballast and the controller is not properly connected. |
M64 |
LAMPS ON: SHOULD BE OFF |
Xenon lamp is on: it should be off. The Q-Sun is in a dark cycle, but the controller sensor detects light intensity greater than 0.05W/m². Replace the lamp relay. |
M67 |
LAMP RELAY STUCK ON |
The lamp relay is stuck in the engaged position. During "relay diagnosis", the light intensity of the lamp is greater than 0.05W/m². There is a problem with the lamp relay. Replace the relay. |
M68 |
RUN RELAY STUCK ON |
The power relay is stuck in the engaged position. During the "relay diagnosis", the light intensity of the lamp is greater than 0.05W/m². The power relay is broken. Replace the relay. |
M69 |
RELAY CHECK IN PROCESS |
Relay diagnosis in progress. This message is displayed during the M67 or M68 diagnostic process and serves only as an indication. |
M70 |
BLACK PANEL TEMP SENSOR FAIL |
The blackboard thermometer is faulty. The blackboard temperature is below 3°C. The blackboard temperature sensor is broken; replace it. |
M72 |
AIR TEMP SENSOR FAIL |
The air temperature sensor is faulty. The air temperature is below 3°C. The air temperature sensor is broken; replace it. |
M74 |
LAB TEMPERATURE SENSOR FAIL |
The laboratory temperature sensor is faulty. The laboratory temperature is below 3°C. The laboratory temperature sensor is broken; replace it. |
M76 |
HUMIDIFIER TEMP SENSOR FAIL |
The humidifier temperature sensor has a fault. The temperature of the humidifier temperature sensor is below 3°C. Replace the humidifier temperature sensor. |
M77 |
RH SENSOR FAILURE |
The relative humidity sensor is faulty. The reading of the relative humidity sensor is less than 3%. Replace the sensor. |
M80 |
FLASH MEMORY FAILURE |
Program verification error. Consult the Customer Service Department of Q-LAB Corporation. |
M81 |
FLASH DATA CORRUPT |
Parameter and system data errors. Contact the Customer Service Department of Q-LAB Corporation. |
M82 |
RAM CORRUPTED, RAM RELOADED |
Memory data error, data needs to be reinstalled. Contact the Customer Service Department of Q-LAB Corporation. |
The above information has detailed step-by-step instructions and guidance in the random product manual. For the product manual and other assistance, please contact us!