Environmental and Water Quality Requirements for Q-SUN Equipment
1.Laboratory Environmental Temperature and Humidity
● The recommended operating environmental temperature and relative humidity for the equipment are: 23°C ±5°C and 50±25% RH.
● Operating the equipment outside the recommended range (or in rare cases, even within the recommended range) may cause certain standards or test conditions to fail to be met.
● Operation outside the recommended range may lead to chamber temperature and/or humidity failures in the equipment.
● Never operate the equipment under conditions where the laboratory temperature >40°C or relative humidity >80%.
● A dry, clean room free of dust, particles, exhaust gases, or salt mist.
● A room with a heating/ventilation/air conditioning (HVAC) system.
● A location away from windows or vents.
For specific information on achievable temperature/humidity values under different laboratory conditions, please consult Luo Zhong Technology.
2.Water Quality Requirements
Model |
Xe-1-B Xe-1-S |
Xe-1-BC Xe-1-SC |
XE-2-H XE-2-HS |
Xe-3-H Xe-3-HDS XE-3-HS Xe-3-HBS |
XE-3-HC XE-3-HSC |
Inlet water pressure |
Type H: 0.7-6.2 bar (10-90 psi); Type S: 2.1-6.2 bar (30-90 psi) |
||||
Water purity |
Type H: Soluble substances < 20 ppm;Type S: Soluble substances < 1 ppm and suspended silicon < 0.1 ppm |
||||
Water consumption |
0.1L/min(Spray) |
0.1L/min(Spray) |
0.5L/min(Spray) 8L/day(Humidifier) |
0.1L/min(Spray) 0.4L/min(Front spray and back spray) 44L/day(Humidifier) |
0.1L/min(Spray) 44L/day(Humidifier) |
Detailed step-by-step instructions and guidance for the above operations are provided in the product manual included with the equipment. For the product manual or any other assistance, please contact us!